Six Tumor Markers (TM6)
product description
CEA, CYFRA21-1, NSE, CA125, proGRP, SCCA six tumor markers. Detecting the levels of tumor markers in human serum, primarily used for the differentiation diagnosis and staging of tumors, prognosis assessment, treatment monitoring, and recurrence monitoring.
Product Performance:
■ Accuracy: The bias value is within ±10% of labeled value using international reference, national reference and enterprise reference as sample. The recovery range is 85~115%;
■ Lowest detectable limit:
CEA     NSE     CA125     CYFRA21-1    proGRP    SCCA
ng/mL  ng/mL    U/mL       ng/mL             pg/mL       ng/mL
  0.5        0.5        8               0.8                10              0.5
■ The liner range of six detection target:
   CEA      NSE    CA125    CYFRA21-1   proGRP   SCCA
  ng/mL    ng/mL    U/mL          ng/mL        pg/mL      ng/mL
0.5-1000 0.5-200  8-1000       0.8-120      10-5000     0.5-70
■ Repeatability:CV≤ 8.0%.